Skip to content

X-ray scattering illuminates a sample with photons whose wavelength is comparable to an interatomic spacing, then records where the scattered photons go, how much energy they lose, and sometimes how their polarization changes. Far from an absorption edge, the leading elastic signal maps electron-density-weighted crystal structure. Near an edge, virtual core excitation makes the amplitude element, valence, orbital, and polarization selective. Coherent beams add speckle and phase-retrieval information, while energy analysis exposes phonons and electronic collective excitations.

Those capabilities do not make a detector image a direct picture of atoms or charge order. The measured counts combine the sample amplitude with flux, polarization, geometry, absorption, fluorescence, detector response, reciprocal-space acceptance, and resolution. In coherent diffraction, the detector ordinarily loses the scattered phase. Near resonance, the same edge that enhances contrast also changes attenuation and can make the scattering amplitude tensorial.

For example, an extra diffraction peak locates an additional spatial periodicity. Its intensity, polarization dependence, and energy dependence help distinguish atomic displacements from changes in electronic charge or orbital states. The following sections develop those distinctions from the scattering amplitude.

Required background. Reciprocal Lattice supplies diffraction geometry and reciprocal-vector indexing. Structure Factors supplies the relation between spatial or dynamical correlations and momentum-resolved intensity. How Quantum Matter Is Measured supplies calibration, resolution and the distinction between measured intensity and an inferred material structure.

X-ray scattering from charge and structure

Section titled “X-ray scattering from charge and structure”

X-rays primarily couple to electrons. Spatial interference tells us about structure; resolving the outgoing photon energy adds information about dynamics. We begin with weak, single scattering, where the outgoing wave is the sum of contributions from the illuminated sample. Strong multiple scattering in a nearly perfect crystal requires dynamical diffraction theory.

The historical experiments are described in X-Ray Experiments. Here the emphasis is how to interpret modern material measurements.

For a photon in vacuum,

∣k∣=2πλ=Eℏc.\lvert\mathbf k\rvert = \frac{2\pi}{\lambda} = \frac{E}{\hbar c}.

A convenient conversion is

λ[nm]≃1.23984E[keV].\lambda[\mathrm{nm}] \simeq \frac{1.23984}{E[\mathrm{keV}]}.

Thus a 10 keV10\ \mathrm{keV} hard X-ray has λ≃0.124 nm\lambda\simeq0.124\ \mathrm{nm}, comparable with crystal spacings. Soft X-rays near transition-metal LL edges have longer wavelengths and lower penetration depths, but offer direct dipole access to valence-sensitive intermediate states.

Adopt the target-transfer convention

Q=ki−kf,ℏω=Ei−Ef.\mathbf Q = \mathbf k_i-\mathbf k_f, \qquad \hbar\omega = E_i-E_f.

Here Q\mathbf Q is a wavevector transfer; the physical momentum delivered to the sample is ℏQ\hbar\mathbf Q. Positive ω\omega means that the sample gains energy. If 2θ2\theta is the angle between incident and outgoing beams, elastic scattering has

Q=4πλsin⁡θ.Q = \frac{4\pi}{\lambda} \sin\theta.

The accessible reciprocal-space region is therefore set by photon energy, scattering angle, sample orientation, and mechanical shadowing. A nominally accessible reflection can still be blocked by a cryostat, pressure cell, beamstop, detector gap, or polarization geometry.

For a free electron initially at rest, neglecting recoil at photon energies Ei≪mec2E_i\ll m_ec^2, the Thomson cross section for specified incoming and outgoing polarizations is

dσTdΩ=re2∣ϵf∗⋅ϵi∣2,\frac{d\sigma_{\mathrm T}}{d\Omega} = r_e^2 \left| \boldsymbol\epsilon_f^* \cdot \boldsymbol\epsilon_i \right|^2,

where

re=e24πϵ0mec2r_e = \frac{e^2}{4\pi\epsilon_0m_ec^2}

is the classical electron radius. The polarization factor is part of the forward model, not a cosmetic correction. Synchrotron radiation is usually strongly polarized, and an analyzer can separate outgoing polarization channels. Even without an analyzer, the detector sums only over the outgoing states admitted by its geometry.

For an isolated atom with electron number density ρe(r)\rho_e(\mathbf r), the nonresonant amplitude in electron units is

f0(Q)=∫d3r ρe(r)eiQ⋅r,f0(0)=Ne.f_0(\mathbf Q) = \int d^3r\, \rho_e(\mathbf r) e^{i\mathbf Q\cdot\mathbf r}, \qquad f_0(\mathbf 0)=N_e.

The forward amplitude counts the electrons: Ne=ZN_e=Z for a neutral atom and Ne=Z−qN_e=Z-q for an ion of charge +qe+qe. Because the electron cloud has finite extent, f0(Q)f_0(Q) generally decreases with QQ. X-ray diffraction therefore weights heavy atoms strongly at small momentum transfer and can have weak contrast for light atoms beside heavy ones. This contrasts with neutron scattering lengths, which vary irregularly across elements and isotopes.

The Fourier integral permits an anisotropic electron density; only a spherical density reduces it to a function of QQ alone. Treating a crystal as a sum of such isolated atoms is an additional approximation. Bonding charge, magnetization, and resonant transitions can require more detailed amplitudes. The IUCr structure-factor introduction explains the connection between atomic positions and interference.

Crystal Structure from Elastic Diffraction

Section titled “Crystal Structure from Elastic Diffraction”

Let Rn\mathbf R_n label unit cells and rj\mathbf r_j the position of atom jj inside a cell. In the kinematic approximation, the elastic amplitude separates into

A(Q,E)=∑neiQ⋅RnFcell(Q,E),\mathcal A(\mathbf Q,E) = \sum_n e^{i\mathbf Q\cdot\mathbf R_n} F_{\mathrm{cell}}(\mathbf Q,E),

with

Fcell(Q,E)=∑joj fj(Q,E) e−Wj(Q)×eiQ⋅rj.F_{\mathrm{cell}}(\mathbf Q,E) ={} \sum_j o_j\, f_j(\mathbf Q,E)\, e^{-W_j(\mathbf Q)} \times e^{i\mathbf Q\cdot\mathbf r_j}.

Here ojo_j is occupancy, fjf_j is the atomic amplitude (with any tensor contracted with the selected incoming and outgoing polarizations), and e−Wje^{-W_j} is the amplitude Debye–Waller factor. For a Gaussian displacement tensor UjU_j,

Wj(Q)=12QTUjQ.W_j(\mathbf Q) = \frac{1}{2} \mathbf Q^{\mathsf T} U_j \mathbf Q.

The lattice sum concentrates an ideal infinite-crystal signal at reciprocal vectors G\mathbf G. The unit-cell amplitude decides the relative intensity and systematic absences. A reciprocal-lattice point is therefore a possible Bragg position, not a guarantee of nonzero intensity.

This formula also shows why structure determination is an inverse problem. Positions, occupancies, displacement parameters, atomic form factors, scale, extinction, preferred orientation, and domain fractions can be correlated. A low residual does not prove that every parameter is identifiable.

Bragg positions and integrated intensities

Section titled “Bragg positions and integrated intensities”

At an elastic Bragg condition,

Q=Ghkl,\mathbf Q=\mathbf G_{hkl},

with dhkl=2π/∣Ghkl∣d_{hkl}=2\pi/|\mathbf G_{hkl}|, so

2dhklsin⁡θ=λ.2d_{hkl}\sin\theta=\lambda.

The alternative notation 2dsin⁡θ=nλ2d\sin\theta=n\lambda holds the plane family fixed and labels its harmonics by nn. That harmonic has Q=nGhkl=Gnh,nk,nl\mathbf Q=n\mathbf G_{hkl}=\mathbf G_{nh,nk,nl}; the order is already included when the reflection is indexed by its full Miller indices.

Peak positions constrain the lattice metric and orientation. Integrated intensities constrain the motif. Peak shapes constrain finite size, strain distributions, defects, mosaicity, correlations, and instrumental resolution. These statements are cleaner than saying that a diffraction pattern simply “shows the structure.”

A schematic integrated-intensity model is

Ihklobs=C Φ t mhkl Lhkl Phkl Ahkl Xhkl×∣Fhkl∣2+Bhkl,I_{hkl}^{\mathrm{obs}} ={} C\, \Phi\,t\, m_{hkl}\, L_{hkl}\, P_{hkl}\, A_{hkl}\, X_{hkl} \times \left| F_{hkl} \right|^2 +B_{hkl},

where Φ\Phi is incident flux, tt is exposure, mm is the multiplicity when symmetry-equivalent powder reflections are combined, LL and PP are geometry and polarization factors, AA is absorption, XX represents extinction or other sample-dependent corrections, and BB is background. The exact factors depend on powder, single-crystal, grazing-incidence, reflectivity, and area-detector geometry.

Correction order matters. A pixel mask and dark subtraction precede geometric integration; monitor normalization must use a linear incident-flux record; dead-time correction must precede comparisons across count rate. Applying a software “Lorentz-polarization correction” without stating the scan geometry is not reproducible.

Powder, single-crystal, and thin-film measurements

Section titled “Powder, single-crystal, and thin-film measurements”

Powder diffraction collapses three-dimensional reciprocal-space shells onto QQ or 2θ2\theta. It is efficient for phase identification and lattice parameters but creates peak overlap, preferred-orientation sensitivity, and limited access to diffuse anisotropy. Rietveld refinement compares the full pattern with a structural and instrumental model; it does not turn each visible peak into an independently measured structure factor.

Single-crystal diffraction retains directional information. Rocking, rotation, and detector scans sample a three-dimensional reciprocal-space volume. Orientation matrices, detector distance and tilt, wavelength, goniometer offsets, and sample displacement must be calibrated together. Symmetry-equivalent reflections test corrections and crystal quality.

Thin films and interfaces introduce truncation rods, finite-thickness fringes, substrate peaks, refraction near grazing incidence, footprint variation, and multiple orientation domains. A broad feature may reflect thickness or mosaic spread rather than a short intrinsic correlation length. Reciprocal-space maps should preserve both in-plane and out-of-plane coordinates before any one-dimensional cut is interpreted.

An ordinary detector records intensity,

Ihkl∝∣Fhkl∣2,I_{hkl} \propto \left| F_{hkl} \right|^2,

not the complex phase of FhklF_{hkl}. For nonresonant electron-density structure factors, an inverse Fourier series would require both:

ρ(r)=1Vcell∑GFGe−iG⋅r.\rho(\mathbf r) = \frac{1}{V_{\mathrm{cell}}} \sum_{\mathbf G} F_{\mathbf G} e^{-i\mathbf G\cdot\mathbf r}.

A complex resonant tensor amplitude is not simply the Fourier coefficient of the ground-state electron density, so this inversion cannot be applied to it without a scattering model. Crystallographic phase information is supplied indirectly by symmetry, chemical constraints, direct or dual-space methods, anomalous contrast, molecular replacement in suitable settings, or refinement from an initial model. A Fourier map reconstructed from model phases is not independent confirmation of that model.

A defensible refinement reports the structural model, scattering-factor tables, constraints and restraints, excluded regions, absorption and extinction treatment, goodness-of-fit measures, parameter covariance or uncertainty, residual density, and tests against plausible alternative space groups or domain models.

Peak Widths, Diffuse Scattering, and Correlations

Section titled “Peak Widths, Diffuse Scattering, and Correlations”

Finite correlation length replaces an ideal reciprocal-space delta peak by a profile. For an exponentially decaying one-dimensional correlation,

C(x)∝e−∣x∣/ξcos⁡(Q0x),C(x) \propto e^{-\lvert x\rvert/\xi} \cos(Q_0x),

the Fourier transform contains two Lorentzians, centered at Q=±Q0Q=\pm Q_0:

C~(Q)∝ξ1+ξ2(Q−Q0)2+ξ1+ξ2(Q+Q0)2.\widetilde C(Q)\propto{}\frac{\xi}{1+\xi^2(Q-Q_0)^2} +\frac{\xi}{1+\xi^2(Q+Q_0)^2}.

When Q0ξ≫1Q_0\xi\gg1, the peaks are well separated. Near +Q0+Q_0 one then obtains the familiar isolated line

I(Q)∝1(Q−Q0)2+ξ−2.I(Q) \propto \frac{1}{ (Q-Q_0)^2+\xi^{-2} }.

After separating the opposite satellite and any background, its intrinsic half width at half maximum is κ=ξ−1\kappa=\xi^{-1}. This conversion is model and convention dependent. A Lorentzian-squared profile, finite domain, algebraic correlation, anisotropic disorder, or distribution of lattice constants gives a different relation. Instrument resolution must be convolved with the intrinsic profile before ξ\xi is reported.

Diffuse scattering between Bragg peaks can reveal short-range order, correlated displacements, stacking faults, chemical disorder, thermal motion, and critical fluctuations. It is often weak compared with air, windows, fluorescence, Compton scattering, substrate, and detector backgrounds. Subtracting a high-temperature pattern can isolate a changing component, but it also removes everything that is temperature independent and mixes thermal-expansion shifts into derivative-like residuals.

Integrated intensity, peak height, and width carry different information. If a peak narrows while conserving integrated weight, its maximum grows without a corresponding growth of total correlation strength. Report the fitted profile and integrated area, not only the brightest pixel.

Near a core-level absorption edge, the scalar anomalous form is often written

f(Q,E)=f0(Q)+f′(E)+if′′(E).f(\mathbf Q,E) = f_0(\mathbf Q) +f'(E) +if''(E).

f′′f'' is tied to absorption and f′f' to it through a dispersion relation. Tabulated free-atom values are useful far from sharp chemical structure, but a resonant experiment deliberately works where oxidation state, ligand field, hybridization, symmetry, and polarization can invalidate a purely scalar table.

A more faithful electric-dipole amplitude is a tensor:

Ares∝ϵf∗⋅f(E,Q)⋅ϵi.\mathcal A_{\mathrm{res}} \propto \boldsymbol\epsilon_f^* \cdot \mathbf f(E,\mathbf Q) \cdot \boldsymbol\epsilon_i.

The tensor contains amplitudes for exciting and de-exciting core states, with their polarization-dependent matrix elements and finite lifetimes. The Kramers–Heisenberg expression below displays those intermediate states explicitly. Higher multipoles can matter where electric-dipole transitions are weak or forbidden. When combining a resonant calculation with a structural amplitude, use consistent phase and scattering-factor conventions; relative phases determine their interference.

The tensor character is an opportunity and a warning. Incident energy, incident and outgoing polarization, azimuthal angle, scattering plane, and crystal orientation can distinguish scalar charge, anisotropic orbital, and magnetic channels. The same dependencies also mean that a single energy scan at one geometry rarely determines a unique microscopic tensor.

The incident and outgoing waves both attenuate inside the sample. Consider a flat uniform slab of normal thickness tt, with both beams crossing the same surface. Let αi,αf\alpha_i,\alpha_f be their positive angles to that surface and μi,μf\mu_i,\mu_f their intensity attenuation coefficients. A scattering event at normal depth zz receives the intensity weight e−gze^{-gz}, where

g=μisin⁡αi+μfsin⁡αf.g=\frac{\mu_i}{\sin\alpha_i}+\frac{\mu_f}{\sin\alpha_f}.

For uniform sources whose contributions from different depths add incoherently, the effective contributing depth is

A(t)=∫0te−gz dz=1−e−gtg.A(t)=\int_0^t e^{-gz}\,dz=\frac{1-e^{-gt}}{g}.

This has units of length. The attenuation relative to the same unattenuated slab is A(t)/tA(t)/t, which tends to one for gt≪1gt\ll1 and to 1/(gt)1/(gt) for gt≫1gt\gg1. Fluorescence is a useful example; its yield also contains the energy-dependent excitation probability and branching fraction, not just this depth factor. Booth and Bridges derive that geometry explicitly.

For a coherently illuminated crystalline slab, add amplitudes first. In a scalar kinematic model, with constant local scattering amplitude and normal mismatch δq\delta q from the Bragg condition,

B(t,δq)=∫0te−gz/2eiδqz dz=1−e−(g/2−iδq)tg/2−iδq,Icoh(δq)∝∣B(t,δq)∣2.\begin{aligned} B(t,\delta q)&=\int_0^t e^{-gz/2}e^{i\delta q z}\,dz\\ &=\frac{1-e^{-(g/2-i\delta q)t}}{g/2-i\delta q},\\ I_{\rm coh}(\delta q)&\propto |B(t,\delta q)|^2. \end{aligned}

The factor 1/21/2 appears because field amplitudes decay as the square root of intensity. At exact Bragg alignment the attenuation relative to an unabsorbed peak is

Icoh(0)Icoh(0)(0)=[1−e−gt/2gt/2]2.\frac{I_{\rm coh}(0)}{I_{\rm coh}^{(0)}(0)} =\left[\frac{1-e^{-gt/2}}{gt/2}\right]^2.

This is a peak intensity, not a universal correction for integrated diffraction. Integrating the whole idealized profile over δq\delta q gives 2πA(t)2\pi A(t) by Parseval’s identity. The measured scan range and geometry therefore matter. Refraction, dynamical diffraction, layering, finite footprint, and standing waves can invalidate this simple slab calculation.

Calculated coherent slab fringes and different absorption factors for peak intensity and independent depth contributions

Top: ∣B/t∣2|B/t|^2 versus the dimensionless mismatch δqt\delta q t for an unabsorbed slab and for gt=2gt=2. Interference creates the thickness fringes; absorption changes their shape. Bottom: the depth factor A/tA/t and the coherent peak factor ∣B(t,0)/t∣2|B(t,0)/t|^2 both approach one in a thin slab, but differ at larger optical thickness. Each curve is normalized to its own unabsorbed signal. The curves follow the equations above, without instrumental broadening or multiple scattering.

Near an edge, the scattering amplitude and attenuation both change. For elastic photons, Ef=EiE_f=E_i and the homogeneous material has the same absorption coefficient at entry and exit. A fluorescence photon instead has a different, approximately fixed energy while EiE_i is scanned. Comparing angles, outgoing-energy channels, and off-resonance scans helps separate these effects; dividing a diffraction trace by an arbitrary absorption trace does not.

A charge or orbital modulation at wavevector qco\mathbf q_{\mathrm{co}} produces satellite intensity near

Q=G±qco\mathbf Q = \mathbf G \pm \mathbf q_{\mathrm{co}}

for a real single-harmonic modulation, together with symmetry-related positions. The paired positions do not require equal measured intensities: polarization, absorption, and complex resonant scattering factors can distinguish them. Higher harmonics can produce additional satellites. Nonresonant intensity may arise primarily from atomic displacements induced by the order. Near an absorption edge, a spatial modulation of the resonant tensor adds element- and orbital-selective contrast:

FG+qco(E)=Fdisp(E)+ϵf∗⋅Δfqco(E)⋅ϵi.F_{\mathbf G+\mathbf q_{\mathrm{co}}}(E) ={} F_{\mathrm{disp}}(E) + \boldsymbol\epsilon_f^* \cdot \Delta\mathbf f_{\mathbf q_{\mathrm{co}}}(E) \cdot \boldsymbol\epsilon_i.

The two amplitudes interfere. Intensity alone does not generally separate lattice displacement from valence or orbital modulation because

I∝∣Fdisp+Fres∣2I \propto \left| F_{\mathrm{disp}} + F_{\mathrm{res}} \right|^2

contains a cross term and loses the overall phase.

To identify charge order, compare the satellite positions and widths with their temperature dependence and with the predicted energy, polarization, and azimuthal response. Off-resonance diffraction tests accompanying lattice displacements. Symmetry-related scans and a second probe help distinguish a genuine modulation from multiple diffraction, a substrate reflection, or beam damage. The 2012 measurements of Ghiringhelli and collaborators and the review by Comin and Damascelli illustrate how these tests constrain cuprate charge correlations.

Resonant enhancement identifies sensitivity to states associated with the chosen edge. It does not directly count “holes per site” without a calibrated microscopic model. Nor does a finite-width quasielastic peak prove truly static order: the statement is limited by energy resolution and acquisition time.

The phase physics, nesting question, reconstructed bands, collective modes, and competition with superconductivity remain at Charge and Spin Density Waves.

If the illumination is coherent across the relevant sample region, amplitudes from different points interfere with stable relative phase. A finite, nonuniform object produces a speckle pattern rather than a smooth ensemble-averaged intensity. In the far field and single-scattering approximation, for uniform coherent illumination,

I(Q)=∣∫d3r ρ(r)eiQ⋅r∣2.I(\mathbf Q) = \left| \int d^3r\, \rho(\mathbf r) e^{i\mathbf Q\cdot\mathbf r} \right|^2.

This is a three-dimensional Fourier transform sampled on the accessible scattering surface. A two-dimensional projection approximation additionally neglects longitudinal phase variation and replaces ρ(r)\rho(\mathbf r) by its density integrated along the incident beam. The detector still records only the modulus squared. Coherent diffraction imaging uses oversampling plus constraints such as finite support, positivity where justified, measured amplitudes, overlap, or known reference structure to recover a compatible complex object iteratively.

Oversampling supplies additional constraints for the common isolated-object reconstruction problem, but it does not guarantee a unique or correct reconstruction; the requirements change when known references or overlapping probe positions supply information. Missing central pixels behind a beamstop, saturation, partial coherence, background, finite detector range, position jitter, probe uncertainty, and stagnation of the phase-retrieval algorithm can create families of solutions. Reconstruct from multiple random starts, show convergence metrics, propagate data uncertainty, and test the result by forward diffraction.

Bragg coherent diffraction and ptychography

Section titled “Bragg coherent diffraction and ptychography”

Near a reciprocal vector G\mathbf G, a strained crystalline object can be represented schematically as

OG(r)=ρG(r)eiG⋅u(r).O_{\mathbf G}(\mathbf r) = \rho_{\mathbf G}(\mathbf r) e^{i\mathbf G\cdot\mathbf u(\mathbf r)}.

The reconstructed amplitude reflects shape, composition, and coherent crystalline fraction. Its phase constrains the displacement projection

ϕG(r)=G⋅u(r)(mod2π).\phi_{\mathbf G}(\mathbf r) = \mathbf G\cdot\mathbf u(\mathbf r) \pmod{2\pi}.

The sign here follows the positive Fourier phase eiQ⋅re^{i\mathbf Q\cdot\mathbf r} used above. One reflection gives one displacement projection. An arbitrary constant phase leaves all diffraction intensities unchanged, so an absolute rigid translation cannot be recovered. Changing the reciprocal-space origin adds a linear phase ramp and changes the reference for mean strain. After choosing those references, spatial phase gradients constrain displacement gradients. At least three linearly independent reflections with common spatial registration are needed for the three displacement components; phase wrapping and crystal defects must also be treated. Yang and collaborators discuss practical strain reconstruction.

Ptychography scans an overlapping coherent probe across an extended sample and jointly infers probe and object. Bragg ptychography combines that redundancy with lattice sensitivity. These methods are computational microscopes, but their spatial resolution is supported by reproducible transfer and phase-retrieval tests, not merely by reconstruction pixel size.

Time-dependent speckle correlations form X-ray photon correlation spectroscopy. A common normalized observable is

g2(Q,τ)=⟨I(Q,t)I(Q,t+τ)⟩⟨I(Q,t)⟩2.g_2(\mathbf Q,\tau) = \frac{ \langle I(\mathbf Q,t)I(\mathbf Q,t+\tau)\rangle }{ \langle I(\mathbf Q,t)\rangle^2 }.

For stationary, zero-mean, jointly circular complex Gaussian scattered fields, with no coherent reference mixed into the signal, the Siegert relation gives

g2(Q,τ)=1+β∣g1(Q,τ)∣2,g_2(\mathbf Q,\tau)=1+\beta|g_1(\mathbf Q,\tau)|^2,

where g1g_1 is the normalized field correlation and β\beta describes measured speckle contrast. Relating g1g_1 to a material correlation function requires the scattering model. A coherent static reference introduces heterodyne terms; non-Gaussian fields need not satisfy this relation. Photon self-counting at zero delay and finite exposure need separate treatment. For aging samples the correlations may depend on both observation times, as explained by Bikondoa. Beam-induced motion and detector afterglow can mimic dynamics.

Resolving the outgoing photon energy distinguishes fluctuations by how much energy they exchange with the sample. A phonon produces a small loss or gain; electronic excitations can appear at much larger transfers.

Away from an absorption edge, the nonresonant density coupling gives

d2σdΩ dε=re2kfki∣ϵf∗⋅ϵi∣2Sρ,E(Q,ε),\frac{d^2\sigma}{d\Omega\,d\varepsilon} =r_e^2\frac{k_f}{k_i} \left|\boldsymbol\epsilon_f^*\cdot\boldsymbol\epsilon_i\right|^2 S_{\rho,E}(\mathbf Q,\varepsilon),

where ε=Ei−Ef\varepsilon=E_i-E_f and Sρ,ES_{\rho,E} is the total illuminated-sample electron-density spectrum per unit energy, not per electron. With ρQ=∑ae−iQ⋅ra\rho_{\mathbf Q}=\sum_a e^{-i\mathbf Q\cdot\mathbf r_a}, our positive target-transfer convention uses

Sρ,E(Q,ε)=∑m,npm∣⟨n∣ρ−Q∣m⟩∣2×δ(ε−En+Em).S_{\rho,E}(\mathbf Q,\varepsilon) ={}\sum_{m,n}p_m |\langle n|\rho_{-\mathbf Q}|m\rangle|^2 \times\delta(\varepsilon-\mathcal E_n+\mathcal E_m).

Here pmp_m is the initial-state probability. If a spectrum is instead defined with δ(ω−ωnm)\delta(\omega-\omega_{nm}), then Sρ,E(Q,ε)=Sρ,ω(Q,ε/ℏ)/ℏS_{\rho,E}(\mathbf Q,\varepsilon)=S_{\rho,\omega}(\mathbf Q,\varepsilon/\hbar)/\hbar. A per-particle definition also requires restoring its particle-number factor. Baron’s review develops the cross section with an energy-normalized spectrum; compare Fourier and momentum signs before transferring a formula between references.

For small energy transfers in a crystal, an atom’s electrons largely follow its nucleus, leading to atomic form factors and phonon displacement correlations. High-resolution IXS can measure these modes in small samples and pressure cells. Higher-energy density scattering can reveal plasmons and particle–hole continua. Multiphonon contributions, Compton scattering, and instrumental resolution still shape the measured spectrum. See Structure Factors for correlation-function normalization and sum rules.

Resonant inelastic X-ray scattering, or RIXS, tunes EiE_i near an absorption edge and resolves EfE_f. For an initial pure material state ∣g⟩|g\rangle in the resonant electric-dipole approximation, its Kramers–Heisenberg intensity is schematically

I(Q,ω)∝∑f∣∑n⟨f∣Df†∣n⟩⟨n∣Di∣g⟩Eg+Ei−En+iΓn/2∣2×δ(Eg+Ei−Ef−Ef),I(\mathbf Q,\omega) \propto \sum_f \left| \sum_n \frac{ \langle f|\mathcal D_f^\dagger|n\rangle \langle n|\mathcal D_i|g\rangle }{ \mathcal E_g+E_i-\mathcal E_n+i\Gamma_n/2 } \right|^2 \times \delta \left( \mathcal E_g+E_i-\mathcal E_f-E_f \right),

Here EfE_f is the outgoing photon energy, while Eg\mathcal E_g, En\mathcal E_n, and Ef\mathcal E_f are material-state energies. The ket ∣f⟩|f\rangle labels the final material state. At finite temperature, average these intensities over initial states with their probabilities. The widths Γn\Gamma_n are in energy units. The review by Ament and collaborators derives this expression and the assumptions behind its simpler effective operators.

The intermediate core hole, dipole selection rules, polarization, edge, and interference among intermediate states shape the spectrum. A RIXS intensity is therefore not generally one universal S(Q,ω)S(\mathbf Q,\omega). In controlled limits, effective operators connect particular channels to spin, orbital, charge, or lattice correlation functions. Outside those limits, multiplet, band, cluster, or many-body calculations are part of the inference.

Typical energy-loss features include:

Feature Typical interpretation Essential caution
elastic or quasielastic line static or slow order, diffuse scattering unresolved phonons, drift, and instrumental tails
phonon progression lattice excitation and electron–phonon coupling self-absorption and unresolved modes alter weights
magnon or multimagnon feature spin excitation in a resonantly allowed channel polarization and model-dependent effective operator
crystal-field or dd–dd excitation local orbital or multiplet transition hybridization and site symmetry matter
charge-transfer excitation ligand–metal or interband process broad continua need a material calculation
plasmon or particle-hole continuum collective or incoherent charge dynamics background and channel mixing

A resonant scan with no outgoing-energy analysis integrates all energies admitted by the detector, including unresolved inelastic scattering and fluorescence. Calling the resulting signal elastic requires additional justification. RIXS separates energy loss, at a substantial cost in throughput.

From detector pixels to scattering intensity

Section titled “From detector pixels to scattering intensity”

A laboratory diffractometer can measure lattice parameters, phases, and many structural changes with a characteristic tube line. Synchrotron sources add tunable energy, controlled polarization, focused beams, and coherent flux. The relevant quantities are the measured energy bandwidth, beam size and divergence, polarization purity, and flux at the sample.

An area-detector pixel becomes a scattered ray only after calibrating beam center, sample-to-detector distance, detector rotations, and pixel size. A powder or crystal standard constrains the geometry; the sample orientation maps laboratory wavevectors to reciprocal coordinates. Inspect calibration residuals across the detector: matching one peak does not ensure that a whole reciprocal-space map is accurate.

Independent ideal photon counts have Poisson variance equal to their mean. Dark subtraction, normalization by a fluctuating monitor, interpolation, and pixel merging change the uncertainty and can create correlations. Detector dead time, saturation, charge sharing, and drift require additional checks. Repeated short exposures help distinguish these effects from photon counting noise.

For incident flux density Φ\Phi and exposure time TT, a useful expected-count model is

d‾i=ΦT∫dΩ dEf Ri(Ω,Ef)×η(Ef)d2σsampledΩ dEf+bi.\overline d_i={}\Phi T\int d\Omega\,dE_f\,R_i(\Omega,E_f) \times\eta(E_f)\frac{d^2\sigma_{\rm sample}}{d\Omega\,dE_f}+b_i.

The sample cross section includes the illuminated material and its attenuation; RiR_i is the dimensionless probability that an outgoing photon reaches bin ii, including resolution and acceptance, η\eta is efficiency, and bib_i is expected background. This expression has the units of counts. Average it over the incident spectrum or time-varying beam when necessary. Elastic scattering contributes a delta function in EfE_f.

A map in (Q,ε)(\mathbf Q,\varepsilon) is a change of coordinates with an associated measure and Jacobian. A fixed incident beam and energy do not sample an arbitrary four-dimensional d3Q dεd^3Q\,d\varepsilon volume; rotations and energy scans provide additional coverage.

Angular divergence, beam bandwidth, footprint, sample mosaic, scan step, detector point-spread, analyzer acceptance, and energy bandwidth produce a coupled resolution volume. A one-dimensional Gaussian width subtracted in quadrature is justified only when both intrinsic and resolution profiles are Gaussian in the fitted coordinate and cross-coordinate coupling is negligible.

For weak or resolution-limited peaks, fit the intrinsic model after convolution with a measured or simulated resolution function. Report upper or lower bounds when the data do not identify both intrinsic width and amplitude.

Air, windows, substrate, sample mount, fluorescence, Compton scattering, and neighboring Bragg tails can all contribute. Multiple diffraction and harmonic contamination may create sharp peaks, while detector gaps or hot pixels create instrumental features.

An off-sample region may not reproduce sample fluorescence or absorption. An off-resonance scan changes both contrast and attenuation. A high-temperature scan may contain real critical or structural scattering. Treat each subtraction as a measured control with its own exposure, normalization, and uncertainty.

X-rays can heat, ionize, charge, reduce, oxidize, desorb, create defects, move domain walls, or alter strain. The absence of visible surface damage is not a sufficient test. Repeat short scans at fresh positions, vary dose rate and total dose independently, monitor a stable structural reflection, and record time order. A signal that evolves monotonically with exposure is not an equilibrium temperature trend.

Mode Primary data product Strong inference Persistent ambiguity
powder diffraction intensity versus 2θ2\theta or QQ phases, lattice metrics, average structure overlap, texture, minority phases
single-crystal diffraction three-dimensional Bragg intensities symmetry, motif, domains, displacement parameters phase problem, extinction, model covariance
diffuse scattering intensity between Bragg positions short-range correlations and disorder models nonunique real-space inversion
resonant elastic scattering Q\mathbf Q, EiE_i, polarization, and azimuth dependence element- and symmetry-selective order absorption and tensor-model dependence
coherent diffraction oversampled speckles constrained complex object or correlation dynamics phase-retrieval nonuniqueness and partial coherence
nonresonant IXS momentum- and energy-resolved density scattering phonons and charge dynamics weak cross section and resolution convolution
RIXS resonant momentum-, loss-, and polarization-resolved intensity selected spin, orbital, charge, and lattice excitations intermediate-state and effective-operator dependence

The modes are complementary, not interchangeable. A resonant satellite can identify an electronic symmetry channel but may provide a less direct displacement amplitude than hard-X-ray diffraction. Coherent imaging can localize strain but only in the illuminated coherent volume and under reconstruction constraints. RIXS can access excitations in tiny samples but may not provide absolute sum-rule normalization comparable with a simpler nonresonant channel.

Ask three questions before assigning a new phase. Where is the peak? Its wavevector and symmetry distinguish candidate modulations. What carries its intensity? The structure factor, polarization and resonant contrast determine which motions or electronic states are visible. How sharp and slow is it? The measured width and energy window constrain correlation lengths and times only after accounting for resolution. These questions connect a detector image to a material model without treating the image as a direct picture of atoms.

1. Photon wavelength and wavevector transfer

Section titled “1. Photon wavelength and wavevector transfer”

A hard-X-ray experiment uses Ei=10.0 keVE_i=10.0\ \mathrm{keV}. Find λ\lambda, kk, and the elastic wavevector transfer at scattering angle 2θ=60∘2\theta=60^\circ. Use hc=1.23984 keV nmhc=1.23984\ \mathrm{keV\,nm}.

Solution

The wavelength is

λ=1.23984 keV nm10.0 keV=0.123984 nm.\lambda = \frac{1.23984\ \mathrm{keV\,nm}}{10.0\ \mathrm{keV}} = 0.123984\ \mathrm{nm}.

Therefore

k=2πλ≃50.68 nm−1=5.068 A˚−1.k = \frac{2\pi}{\lambda} \simeq 50.68\ \mathrm{nm}^{-1} = 5.068\ \mathrm{\mathring A}^{-1}.

Because θ=30∘\theta=30^\circ,

Q=2ksin⁡θ=k≃5.068 A˚−1.Q = 2k\sin\theta = k \simeq 5.068\ \mathrm{\mathring A}^{-1}.

This is a kinematic result. Mechanical access and polarization can still make that point unusable.

A conventional cubic cell contains two identical atoms at fractional coordinates (0,0,0)(0,0,0) and (1/2,1/2,1/2)(1/2,1/2,1/2). Neglect displacement factors. Find the elastic structure factor and the extinction rule.

Solution

For Miller indices (hkl)(hkl),

Fhkl=f[1+e2πi(h+k+l)/2]=f[1+(−1)h+k+l].\begin{aligned} F_{hkl} &= f \left[ 1+ e^{2\pi i(h+k+l)/2} \right] \\ &= f \left[ 1+(-1)^{h+k+l} \right]. \end{aligned}

Thus

Fhkl={2f,h+k+l even,0,h+k+l odd.F_{hkl} = \begin{cases} 2f, & h+k+l\ \text{even},\\ 0, & h+k+l\ \text{odd}. \end{cases}

The integer triples label the conventional cubic reciprocal-space grid. With identical atoms at the origin and body center, the actual Bravais lattice is body-centered cubic; odd-sum triples are absent from its reciprocal lattice. Describing the same crystal with a conventional cubic cell and a two-site motif reproduces those absences through the structure factor. The indexing grid should not be confused with the crystal’s physical reciprocal lattice.

Two sublattices contribute with opposite phase at a superlattice reflection. Far from an edge they have equal form factor f0f_0. Near the edge, sublattice A changes to f0+δf(E)f_0+\delta f(E) while B remains f0f_0. What is the superlattice amplitude and what does the measured intensity determine?

Solution

The opposite phase gives

Fsup(E)=[f0+δf(E)]−f0=δf(E).F_{\mathrm{sup}}(E) = \left[ f_0+\delta f(E) \right] -f_0 = \delta f(E).

The ideal intensity is

Isup(E)∝∣δf(E)∣2.I_{\mathrm{sup}}(E) \propto \left| \delta f(E) \right|^2.

The resonance makes the sublattice difference visible, but intensity alone does not determine the complex phase of δf\delta f, nor does it convert directly to a charge difference. Absorption, polarization, other structural amplitudes, and a microscopic scattering model are needed in a real experiment.

A Lorentzian satellite has observed half width at half maximum κobs=0.020 A˚−1\kappa_{\mathrm{obs}}=0.020\ \mathrm{\mathring A}^{-1}. A reference reflection shows a Lorentzian instrumental half width κR=0.006 A˚−1\kappa_R=0.006\ \mathrm{\mathring A}^{-1} in the same scan coordinate. Assuming an isolated satellite, Lorentzian profiles in this scan coordinate, and exponential intrinsic correlations, estimate ξ\xi.

Solution

Convolution adds Lorentzian half widths:

κobs=κint+κR.\kappa_{\mathrm{obs}} = \kappa_{\mathrm{int}} + \kappa_R.

Hence

κint=0.020−0.006=0.014 A˚−1.\kappa_{\mathrm{int}} = 0.020-0.006 = 0.014\ \mathrm{\mathring A}^{-1}.

For the stated exponential model,

ξ=κint−1≃71 A˚.\xi = \kappa_{\mathrm{int}}^{-1} \simeq 71\ \mathrm{\mathring A}.

Quadrature subtraction would be appropriate for Gaussian widths, not Lorentzian widths. The answer is conditional on the line-shape and one-dimensional resolution assumptions.

5. Independent events and coherent depth interference

Section titled “5. Independent events and coherent depth interference”

A thick flat sample is measured in reflection with αi=αf=30∘\alpha_i=\alpha_f=30^\circ. In an incident-energy scan, μi\mu_i changes from 0.500.50 to 1.00 μm−11.00\ \mathrm{\mu m}^{-1}. The selected fluorescence line has fixed outgoing energy and μf=0.40 μm−1\mu_f=0.40\ \mathrm{\mu m}^{-1}. Holding its local excitation probability artificially fixed to isolate attenuation, find the intensity ratio. Then consider elastic coherent scattering at exact Bragg alignment, where both attenuation coefficients change from 0.500.50 to 1.00 μm−11.00\ \mathrm{\mu m}^{-1}. What ratio follows from the thick-slab kinematic model if the intrinsic amplitude is fixed?

Solution

For independent fluorescence events, I∝A(∞)=1/gI\propto A(\infty)=1/g. Thus

goff=0.50+0.400.5=1.80 μm−1,gon=1.00+0.400.5=2.80 μm−1,Ion/Ioff=1.80/2.80≃0.643.\begin{aligned} g_{\rm off}&=\frac{0.50+0.40}{0.5}=1.80\ \mathrm{\mu m}^{-1},\\ g_{\rm on}&=\frac{1.00+0.40}{0.5}=2.80\ \mathrm{\mu m}^{-1},\\ I_{\rm on}/I_{\rm off}&=1.80/2.80\simeq0.643. \end{aligned}

A real fluorescence excitation probability generally changes near the edge, so 0.6430.643 is only the attenuation contribution.

For the elastic coherent case, goff=2.00 μm−1g_{\rm off}=2.00\ \mathrm{\mu m}^{-1} and gon=4.00 μm−1g_{\rm on}=4.00\ \mathrm{\mu m}^{-1}. At exact Bragg alignment, B(∞,0)=2/gB(\infty,0)=2/g, giving the peak ratio

Ion(0)/Ioff(0)=(2/4)2=14.I_{\rm on}(0)/I_{\rm off}(0)=(2/4)^2=\frac14.

Integrating the whole idealized coherent profile instead gives a ratio goff/gon=1/2g_{\rm off}/g_{\rm on}=1/2. The examples differ in both outgoing photon energy and whether a peak or integrated signal is measured. Keeping only the incident absorption change would be inconsistent for the homogeneous elastic case. These results assume single scattering even though absorption is strong.

A two-dimensional coherent-diffraction experiment measures reconstructed phase differences from a common registered reference point, Δϕ1=0.12\Delta\phi_1=0.12 rad near G1=(G,0)\mathbf G_1=(G,0) and Δϕ2=−0.05\Delta\phi_2=-0.05 rad near G2=(0,G)\mathbf G_2=(0,G) at a second registered point. Assume unwrapped phases and fixed reciprocal-space origins. Express the displacement differences from the reference point in terms of GG. Why would the first reflection alone be insufficient?

Solution

Using ΔϕG=G⋅Δu\Delta\phi_{\mathbf G}=\mathbf G\cdot\Delta\mathbf u,

Δϕ1=GΔux,Δϕ2=GΔuy.\Delta\phi_1 = G\Delta u_x, \qquad \Delta\phi_2 = G\Delta u_y.

Thus

Δux=0.12G,Δuy=−0.05G.\Delta u_x = \frac{0.12}{G}, \qquad \Delta u_y = -\frac{0.05}{G}.

The first reflection constrains only Δux\Delta u_x and is blind to Δuy\Delta u_y. A constant unknown phase for either reflection cancels in its phase difference. A linear phase ramp does not cancel, which is why the reciprocal origins and strain reference must be fixed. Neither reflection determines an out-of-plane component.

  • J. Als-Nielsen and D. McMorrow, Elements of Modern X-Ray Physics, 2nd ed. (Wiley, 2011), doi:10.1002/9781119998365.
  • L. J. P. Ament, M. van Veenendaal, T. P. Devereaux, J. P. Hill, and J. van den Brink, “Resonant Inelastic X-Ray Scattering Studies of Elementary Excitations,” Reviews of Modern Physics 83, 705–767 (2011), doi:10.1103/RevModPhys.83.705.
  • A. Q. R. Baron, “Introduction to High-Resolution Inelastic X-Ray Scattering,” in Synchrotron Light Sources and Free-Electron Lasers (Springer, 2016), doi:10.1007/978-3-319-14394-1_52.
  • O. Bikondoa, “On the Use of Two-Time Correlation Functions for X-Ray Photon Correlation Spectroscopy Data Analysis,” Journal of Applied Crystallography 50, 357–368 (2017), doi:10.1107/S1600576717000577.
  • C. H. Booth and F. Bridges, “Improved Self-Absorption Correction for Fluorescence Measurements of Extended X-Ray Absorption Fine-Structure,” preprint (2003), arXiv:cond-mat/0306252.
  • R. Comin and A. Damascelli, “Resonant X-Ray Scattering Studies of Charge Order in Cuprates,” Annual Review of Condensed Matter Physics 7, 369–405 (2016), doi:10.1146/annurev-conmatphys-031115-011401.
  • D. T. Cromer and D. A. Liberman, “Relativistic Calculation of Anomalous Scattering Factors for X Rays,” Journal of Chemical Physics 53, 1891–1898 (1970), doi:10.1063/1.1674266.
  • B. D. Cullity and S. R. Stock, Elements of X-Ray Diffraction, 3rd ed. (Prentice Hall, 2001).
  • J. R. Fienup, “Phase Retrieval Algorithms: A Comparison,” Applied Optics 21, 2758–2769 (1982), doi:10.1364/AO.21.002758.
  • G. Ghiringhelli et al., “Long-Range Incommensurate Charge Fluctuations in (Y,Nd)Ba2_2Cu3_3O6+x_{6+x},” Science 337, 821–825 (2012), doi:10.1126/science.1223532.
  • J. P. Hannon, G. T. Trammell, M. Blume, and D. Gibbs, “X-Ray Resonance Exchange Scattering,” Physical Review Letters 61, 1245–1248 (1988), doi:10.1103/PhysRevLett.61.1245.
  • B. L. Henke, E. M. Gullikson, and J. C. Davis, “X-Ray Interactions: Photoabsorption, Scattering, Transmission, and Reflection at E=50E=50–30,000 eV30{,}000\ \mathrm{eV}, Z=1Z=1–9292,” Atomic Data and Nuclear Data Tables 54, 181–342 (1993), doi:10.1006/adnd.1993.1013; tabulated data are maintained by CXRO.
  • J. P. Hill and D. F. McMorrow, “X-Ray Resonant Exchange Scattering: Polarization Dependence and Correlation Functions,” Acta Crystallographica A 52, 236–244 (1996), doi:10.1107/S0108767395012670.
  • International Union of Crystallography, “Introduction to the Calculation of Structure Factors”, Teaching Pamphlet No. 3.
  • Joint Committee for Guides in Metrology, Evaluation of Measurement Data — Guide to the Expression of Uncertainty in Measurement, JCGM 100:2008.
  • J. Miao, P. Charalambous, J. Kirz, and D. Sayre, “Extending the Methodology of X-Ray Crystallography to Allow Imaging of Micrometre-Sized Non-Crystalline Specimens,” Nature 400, 342–344 (1999), doi:10.1038/22498.
  • J. Miao, “Computational Microscopy with Coherent Diffractive Imaging and Ptychography,” Nature 637, 281–295 (2025), doi:10.1038/s41586-024-08278-z.
  • M. A. Pfeifer, G. J. Williams, I. A. Vartanyants, R. Harder, and I. K. Robinson, “Three-Dimensional Mapping of a Deformation Field inside a Nanocrystal,” Nature 442, 63–66 (2006), doi:10.1038/nature04867.
  • I. Robinson and R. Harder, “Coherent X-Ray Diffraction Imaging of Strain at the Nanoscale,” Nature Materials 8, 291–298 (2009), doi:10.1038/nmat2400.
  • B. E. Warren, X-Ray Diffraction (Dover, 1990).
  • D. Yang et al., “Refinements for Bragg Coherent X-Ray Diffraction Imaging: Electron Backscatter Diffraction Alignment and Strain Field Computation,” Journal of Applied Crystallography 55, 1184–1195 (2022), doi:10.1107/S1600576722007646.