Skip to content

A linewidth describes the spread of a spectral feature. To interpret it, identify the spectral axis, the definition of width, and the measured quantity. For example, an excited-state lifetime of 3030 ns gives a natural full width of 5.315.31 MHz when the lower state is stable and pure dephasing is absent. The decay rate itself is 3.33×107 s−13.33\times10^7\ \mathrm{s^{-1}}.

This reference collects isolated-line formulas and their assumptions. For the reasoning behind them, see Line Shapes and Broadening.

Helpful background. Line Shapes and Broadening explains how coherence decay and a distribution of resonance frequencies produce different profiles.

Unless a section says otherwise, use angular frequency

x=δω=ω−ω0x = \delta\omega = \omega-\omega_0

and the following symbols:

Symbol Meaning Units on an angular-frequency axis
γL\gamma_{\mathrm L} Lorentzian HWHM rad s−1\mathrm{rad\,s^{-1}}
wL=2γLw_{\mathrm L}=2\gamma_{\mathrm L} Lorentzian FWHM rad s−1\mathrm{rad\,s^{-1}}
σG\sigma_{\mathrm G} Gaussian standard deviation rad s−1\mathrm{rad\,s^{-1}}
wGw_{\mathrm G} Gaussian FWHM rad s−1\mathrm{rad\,s^{-1}}
Γ1\Gamma_1 population-decay rate s−1\mathrm{s^{-1}}
Γ2\Gamma_2 coherence-decay rate s−1\mathrm{s^{-1}}
T1=1/Γ1T_1=1/\Gamma_1 population lifetime s\mathrm s
T2=1/Γ2T_2=1/\Gamma_2 coherence time s\mathrm s

Radians are dimensionless in SI, but retaining rad in a label makes the 2π2\pi convention visible.

Angular frequency ω\omega, ordinary frequency ν\nu, photon energy EE, and spectroscopic wavenumber ν~\widetilde\nu obey

ω=2πν,E=hν=ℏω,ν=cν~.\omega=2\pi\nu, \qquad E=h\nu=\hbar\omega, \qquad \nu=c\widetilde\nu.

These coordinate changes are linear, so corresponding widths satisfy exactly

Δω=2πΔν,\Delta\omega = 2\pi\Delta\nu, ΔE=ℏΔω=hΔν,\Delta E = \hbar\Delta\omega = h\Delta\nu, Δν~=Δνc.\Delta\widetilde\nu = \frac{\Delta\nu}{c}.

Do not use ν~\widetilde\nu for ordinary frequency merely because both are pronounced “nu.” Spectroscopic wavenumber is commonly reported in cm−1\mathrm{cm^{-1}}.

Name Definition Symmetric-profile relation
peak center location of maximum equals the location parameter only for the stated symmetric model
HWHM positive offset at half maximum x1/2x_{1/2}
FWHM distance between the two half-maximum points 2x1/22x_{1/2}
standard deviation square root of second central moment finite for a Gaussian; undefined for an ideal Lorentzian
1/e1/e half width offset where the profile falls to 1/e1/e of peak profile dependent
equivalent width integrated absorption relative to a continuum not a line-profile FWHM

The symbol Γ\Gamma has no universal width meaning. Always replace it with a named HWHM, FWHM, decay rate, or energy width when comparing sources.

Let gx(x)g_x(x) be a line profile normalized on coordinate xx:

∫−∞∞gx(x) dx=1.\int_{-\infty}^{\infty} g_x(x)\,dx = 1.

Its units are the reciprocal units of xx. A spectral power density can be written Px(x)=Ptotgx(x)P_x(x)=P_{\mathrm{tot}}g_x(x), where PtotP_{\mathrm{tot}} is the integrated power. The area stays fixed while the profile redistributes it. Peak height and width are therefore not independent at fixed area.

For a monotonic change y=y(x)y=y(x), a normalized density transforms as

gy(y)=gx ⁣(x(y))∣dxdy∣.g_y(y)=g_x\!\bigl(x(y)\bigr)\left|\frac{dx}{dy}\right|.

For positive vacuum frequencies and wavelengths,

ν=cλ,Pλ(λ)=Pν(c/λ)cλ2.\nu=\frac{c}{\lambda},\qquad P_\lambda(\lambda)=P_\nu(c/\lambda)\frac{c}{\lambda^2}.

The Jacobian ensures that integrating either density over the same spectral interval gives the same power. The ideal profiles below extend detuning to the whole real line; for an optical line this is a local approximation near a positive carrier. Do not apply a wavelength transformation to their unphysical negative-frequency tails.

A monochromatic attenuation coefficient is a different object. In Iout(ν)=Iin(ν)e−α(ν)ℓI_{\rm out}(\nu)=I_{\rm in}(\nu)e^{-\alpha(\nu)\ell}, the coefficient α\alpha is the attenuation per length at a specified frequency. Relabeling the same light by wavelength gives

αλ(λ)=αν(c/λ),\alpha_\lambda(\lambda)=\alpha_\nu(c/\lambda),

with no Jacobian. Both incident and transmitted spectral power densities acquire the same Jacobian, which cancels in their ratio. An integrated absorption strength such as ∫αν dν\int\alpha_\nu\,d\nu consequently depends on the chosen coordinate. In a formula αν=Sνgν\alpha_\nu=S_\nu g_\nu, define SνS_\nu using that same coordinate.

A frequency-symmetric profile generally becomes asymmetric in wavelength. The useful width relation

∣Δλ∣λ0≃∣Δν∣ν0\frac{|\Delta\lambda|}{\lambda_0}\simeq \frac{|\Delta\nu|}{\nu_0}

is only a first-order narrow-line approximation.

Property Lorentzian Gaussian Voigt
parameters center and HWHM γL\gamma_{\mathrm L} center and standard deviation σG\sigma_{\mathrm G} center, γL\gamma_{\mathrm L}, and σG\sigma_{\mathrm G}
common origin exponential coherence, isolated damped pole, impact-limit phase interruption normal distribution of static shifts, thermal one-axis velocities, Gaussian instrument response independent Gaussian and Lorentzian stages
wings algebraic, proportional to x−2x^{-2} exponential in −x2-x^2 Lorentzian asymptote if γL>0\gamma_{\mathrm L}>0
variance does not exist on an infinite domain σG2\sigma_{\mathrm G}^2 does not exist if γL>0\gamma_{\mathrm L}>0
convolution with the same family HWHMs add variances add HWHMs add and Gaussian variances add
one-width description HWHM or FWHM is sufficient after normalization standard deviation or FWHM is sufficient after normalization requires both component widths

Similar-looking cores do not imply similar wings. This matters for weak neighboring lines, optical-depth retrievals, and truncated fit windows.

Equal-width Lorentzian and Gaussian profiles on linear and logarithmic scales

Both curves have unit area and the same full width 2γ2\gamma. Their peak heights differ because the Lorentzian puts more area into its wings. The lower panel shows how rapidly the Gaussian becomes small away from line center. The plotted ordinate is γ\gamma times each normalized profile.

The area-normalized Lorentzian is

L(x;γL)=1πγLx2+γL2.L(x;\gamma_{\mathrm L}) = \frac{1}{\pi} \frac{\gamma_{\mathrm L}} {x^2+\gamma_{\mathrm L}^2}.

Its lookup properties are:

L(0)=1πγL,L(0) = \frac{1}{\pi\gamma_{\mathrm L}}, HWHM=γL,FWHM=wL=2γL,\text{HWHM} = \gamma_{\mathrm L}, \qquad \text{FWHM} = w_{\mathrm L} = 2\gamma_{\mathrm L}, L(x)∼γLπx2(∣x∣≫γL).L(x) \sim \frac{\gamma_{\mathrm L}} {\pi x^2} \qquad (|x|\gg\gamma_{\mathrm L}).

A Lorentzian does not have a finite ordinary variance:

∫−∞∞x2L(x) dxdiverges.\int_{-\infty}^{\infty} x^2L(x)\,dx \quad\text{diverges}.

Consequently, a reported “Lorentzian standard deviation” either refers to a finite window, a different distribution, or an undocumented convention.

An exponentially decaying coherence

C(t)∝e−Γ2∣t∣e−iω0tC(t) \propto e^{-\Gamma_2|t|} e^{-i\omega_0t}

produces a Lorentzian frequency dependence with

γL,ω=Γ2,\gamma_{\mathrm L,\omega} = \Gamma_2,

and hence

wL,ω=2Γ2=2T2.w_{\mathrm L,\omega} = 2\Gamma_2 = \frac{2}{T_2}.

On an ordinary-frequency axis,

γL,ν=Γ22π,\gamma_{\mathrm L,\nu} = \frac{\Gamma_2}{2\pi}, wL,ν=Γ2π=1πT2.w_{\mathrm L,\nu} = \frac{\Gamma_2}{\pi} = \frac{1}{\pi T_2}.

These relations assume exponential coherence over the times relevant to the measurement.

The area-normalized Gaussian is

G(x;σG)=1σG2πexp⁡ ⁣(−x22σG2).G(x;\sigma_{\mathrm G}) = \frac{1} {\sigma_{\mathrm G}\sqrt{2\pi}} \exp\!\left( -\frac{x^2} {2\sigma_{\mathrm G}^2} \right).

Its lookup properties are:

G(0)=1σG2π,G(0) = \frac{1} {\sigma_{\mathrm G}\sqrt{2\pi}}, HWHM=2ln⁡2 σG,\text{HWHM} = \sqrt{2\ln2}\, \sigma_{\mathrm G}, wG=22ln⁡2 σG.w_{\mathrm G} = 2\sqrt{2\ln2}\, \sigma_{\mathrm G}.

Numerically,

wG≃2.35482 σG.w_{\mathrm G} \simeq 2.35482\,\sigma_{\mathrm G}.

The 1/e1/e half width is

x1/e=2 σG.x_{1/e} = \sqrt2\,\sigma_{\mathrm G}.

Do not confuse this with the half width at half maximum.

For independent Gaussian shifts,

σtot2=∑iσi2.\sigma_{\mathrm{tot}}^2 = \sum_i\sigma_i^2.

Because every Gaussian FWHM contains the same multiplicative factor,

wG,tot2=∑iwG,i2.w_{\mathrm{G,tot}}^2 = \sum_i w_{\mathrm{G},i}^2.

This quadrature rule applies to Gaussian convolution parameters. It is not a generic rule for all quoted uncertainties or all spectral widths.

The Voigt profile is the convolution of a Gaussian and a Lorentzian (NIST DLMF, §7.19):

V(x;σG,γL)=∫−∞∞G(x′;σG)×L(x−x′;γL) dx′.V(x;\sigma_{\mathrm G},\gamma_{\mathrm L}) = \int_{-\infty}^{\infty} G(x';\sigma_{\mathrm G}) \times L(x-x';\gamma_{\mathrm L}) \,dx'.

It is area normalized if both components are normalized.

Define

w(z)=e−z2erfc⁡(−iz),w(z) = e^{-z^2} \operatorname{erfc}(-iz),

and

z=x+iγLσG2.z = \frac{x+i\gamma_{\mathrm L}} {\sigma_{\mathrm G}\sqrt2}.

Then

V(x;σG,γL)=Re⁡w(z)σG2π.V(x;\sigma_{\mathrm G},\gamma_{\mathrm L}) = \frac{ \operatorname{Re}w(z) } {\sigma_{\mathrm G}\sqrt{2\pi}}.

Use a tested complex-error-function implementation. Direct numerical quadrature or naive evaluation of e−z2erfc⁡(−iz)e^{-z^2}\operatorname{erfc}(-iz) can lose accuracy in parts of the complex plane.

The Voigt profile has the expected limits:

lim⁡γL→0V(x;σG,γL)=G(x;σG),\lim_{\gamma_{\mathrm L}\to0} V(x;\sigma_{\mathrm G},\gamma_{\mathrm L}) = G(x;\sigma_{\mathrm G}), lim⁡σG→0V(x;σG,γL)=L(x;γL).\lim_{\sigma_{\mathrm G}\to0} V(x;\sigma_{\mathrm G},\gamma_{\mathrm L}) = L(x;\gamma_{\mathrm L}).

Any nonzero γL\gamma_{\mathrm L} controls the far-wing asymptote, even when the core looks nearly Gaussian.

Let

wL=2γL,wG=22ln⁡2 σG.w_{\mathrm L}=2\gamma_{\mathrm L}, \qquad w_{\mathrm G} = 2\sqrt{2\ln2}\,\sigma_{\mathrm G}.

The approximation discussed by Olivero and Longbothum (1977) is

wV≃0.5346 wL+0.2166 wL2+wG2.w_{\mathrm V} \simeq 0.5346\,w_{\mathrm L} + \sqrt{ 0.2166\,w_{\mathrm L}^2 + w_{\mathrm G}^2 }.

This estimates the resulting FWHM. It does not convert a fitted single wVw_{\mathrm V} into unique Lorentzian and Gaussian components.

A pseudo-Voigt profile is a weighted sum,

P(x)=ηL(x)+(1−η)G(x),P(x) = \eta L(x) + (1-\eta)G(x),

with widths arranged according to a chosen approximation. A Voigt profile is a convolution. A pseudo-Voigt can be computationally convenient, but its mixing fraction η\eta is not a physical Lorentzian fraction unless a particular model establishes that interpretation.

Independent stages Correct combination Resulting family
Lorentzian + Lorentzian add HWHMs: γtot=γ1+γ2\gamma_{\mathrm{tot}}=\gamma_1+\gamma_2 Lorentzian
Gaussian + Gaussian add variances: σtot2=σ12+σ22\sigma_{\mathrm{tot}}^2=\sigma_1^2+\sigma_2^2 Gaussian
Lorentzian + Gaussian convolve Voigt
Voigt + Voigt add Lorentzian HWHMs and Gaussian variances separately Voigt
unresolved discrete components sum shifted component profiles with weights generally not one standard profile
saturation plus Doppler distribution average the intensity-dependent response over velocities not automatically a fixed Voigt
line mixing or interference use coupled response amplitudes or a relaxation matrix can be asymmetric or dispersive

Blindly adding all FWHMs linearly is wrong. Blindly adding all FWHMs in quadrature is also wrong. First classify the profile family and statistical relationship.

For a transition between upper and lower levels with population-loss rates Γu\Gamma_u and Γl\Gamma_l, and with additional Markovian pure dephasing γϕ\gamma_\phi, independent population-loss channels give

Γ2=Γu+Γl2+γϕ.\Gamma_2 = \frac{\Gamma_u+\Gamma_l}{2} + \gamma_\phi.

The Lorentzian angular-frequency widths are

γL,ω=Γ2,\gamma_{\mathrm L,\omega} = \Gamma_2, wL,ω=2Γ2=Γu+Γl+2γϕ.w_{\mathrm L,\omega} = 2\Gamma_2 = \Gamma_u+\Gamma_l+2\gamma_\phi.

On an ordinary-frequency axis,

wL,ν=Γu+Γl+2γϕ2π.w_{\mathrm L,\nu} = \frac{ \Gamma_u+\Gamma_l+2\gamma_\phi } {2\pi}.

This assumes an isolated optical coherence with no coupling to other coherences. Shared reservoirs, correlated noise, or nearly degenerate transitions can require a coupled master equation instead.

If the lower state is stable, the upper state decays only radiatively, and there is no pure dephasing,

Γl=0,γϕ=0,\Gamma_l=0, \qquad \gamma_\phi=0, Γu=∑lAul=1τu.\Gamma_u = \sum_l A_{ul} = \frac{1}{\tau_u}.

Then

wnat,ω=Γu,w_{\mathrm{nat},\omega} = \Gamma_u, Δνnat=Γu2π=12πτu,\Delta\nu_{\mathrm{nat}} = \frac{\Gamma_u}{2\pi} = \frac{1}{2\pi\tau_u}, ΔEnat=ℏΓu.\Delta E_{\mathrm{nat}} = \hbar\Gamma_u.

The relation “linewidth equals inverse lifetime” is correct here only for angular-frequency FWHM. The ordinary-frequency FWHM contains 2π2\pi.

The total upper-state loss can include

Γutot=Γurad+Γunr,\Gamma_u^{\mathrm{tot}} = \Gamma_u^{\mathrm{rad}} + \Gamma_u^{\mathrm{nr}},

where nonradiative loss may include quenching, predissociation, autoionization, reaction, or escape from the selected state. Such loss can broaden the transition without contributing to the detected fluorescence branch.

“Natural linewidth” should identify the electromagnetic environment. Cavities and structured reservoirs can modify radiative decay, while strong coupling can replace one broadened peak with resolved normal modes.

For a thermal gas in equilibrium, one velocity component is Gaussian with variance

σv2=kBTm.\sigma_v^2 = \frac{k_{\mathrm B}T}{m}.

For a wavevector of magnitude

k0=ω0c,k_0 = \frac{\omega_0}{c},

a particle moving with velocity component v∥v_\parallel along the beam sees the rest-frame detuning

Δatom≃ωL−k0v∥−ω0.\Delta_{\rm atom}\simeq\omega_{\rm L}-k_0v_\parallel-\omega_0.

Its resonance therefore occurs at the laboratory frequency ωL≃ω0+k0v∥\omega_{\rm L}\simeq\omega_0+k_0v_\parallel. The sign distinguishes a shift of the laser frequency in the particle frame from a shift of the laboratory resonance.

The angular-frequency standard deviation is therefore

σD,ω=ω0ckBTm.\sigma_{\mathrm D,\omega} = \frac{\omega_0}{c} \sqrt{ \frac{k_{\mathrm B}T}{m} }.

The angular-frequency Doppler FWHM is

wD,ω=22ln⁡2 ω0ckBTm.w_{\mathrm D,\omega} = 2\sqrt{2\ln2}\, \frac{\omega_0}{c} \sqrt{ \frac{k_{\mathrm B}T}{m} }.

Equivalently,

ΔνD=2ν02kBTln⁡2mc2,\Delta\nu_{\mathrm D} = 2\nu_0 \sqrt{ \frac{2k_{\mathrm B}T\ln2} {mc^2} },

and

ΔνDν0=8kBTln⁡2mc2.\frac{\Delta\nu_{\mathrm D}}{\nu_0} = \sqrt{ \frac{8k_{\mathrm B}T\ln2} {mc^2} }.

The same fractional formula applies to spectroscopic wavenumber:

Δν~Dν~0=8kBTln⁡2mc2.\frac{ \Delta\widetilde\nu_{\mathrm D} } {\widetilde\nu_0} = \sqrt{ \frac{8k_{\mathrm B}T\ln2} {mc^2} }.

The formula assumes:

  • a Maxwell–Boltzmann distribution;
  • a nonrelativistic first-order Doppler shift;
  • thermal equilibrium;
  • the stated particle or isotopologue mass;
  • the temperature projected along the optical wavevector; and
  • no unresolved structure folded into the fitted Gaussian width.

A Gaussian fit parameter is not automatically a thermodynamic temperature. Laser jitter, isotope structure, field gradients, spatially varying light shifts, and instrumental response can also produce or enlarge a Gaussian component.

In the impact limit, short statistically independent collisions add a homogeneous coherence-decay rate. A useful form is

C(t)∝e−(Γ2(0)+γcol)te−i(ω0+Δcol)t.C(t) \propto e^{- (\Gamma_2^{(0)}+\gamma_{\mathrm{col}})t } e^{-i (\omega_0+\Delta_{\mathrm{col}})t }.

Thus the collision-induced Lorentzian HWHM adds:

γL,tot=γL,0+γcol.\gamma_{\mathrm L,tot} = \gamma_{\mathrm L,0} + \gamma_{\mathrm{col}}.

At low perturber density,

γcol∼np⟨vrelσbroad⟩.\gamma_{\mathrm{col}} \sim n_{\mathrm p} \left\langle v_{\mathrm{rel}} \sigma_{\mathrm{broad}} \right\rangle.

The broadening cross section describes phase interruption. It need not equal an elastic, reactive, or momentum-transfer cross section.

Molecular line lists often tabulate pressure-broadened Lorentzian HWHM coefficients at

Tref=296 K,pref=1 atm.T_{\mathrm{ref}}=296\ \mathrm K, \qquad p_{\mathrm{ref}}=1\ \mathrm{atm}.

With HWHM coefficients bj(Tref)b_j(T_{\rm ref}) expressed per atmosphere and perturber partial pressures pjp_j in atmospheres, a useful form is

γcol(T)=∑jbj(Tref)pj(TrefT)nj.\gamma_{\rm col}(T)=\sum_j b_j(T_{\rm ref})p_j \left(\frac{T_{\rm ref}}{T}\right)^{n_j}.

The result is a wavenumber HWHM if bjb_j is in cm−1 atm−1\mathrm{cm^{-1}\,atm^{-1}}. Each exponent is specific to a transition and perturber. The common HITRAN air/self formula uses pair=p−pselfp_{\rm air}=p-p_{\rm self} and assumes nself=nairn_{\rm self}=n_{\rm air} when no independent self-broadening temperature dependence is available (HITRAN, “Definitions and Units”). Equal exponents are an assumption, not a universal property of collisions. A pressure-independent lifetime width must be added separately.

If tabulated coefficients instead carry width units at a reference pressure, multiply by pj/prefp_j/p_{\rm ref}, not by a dimensional pressure.

A pressure shift is separate:

ν~0(p,T)=ν~0(0)+δ(p,T) p.\widetilde\nu_0(p,T) = \widetilde\nu_0(0) + \delta(p,T)\,p.

Never absorb a fitted center shift into the width parameter.

Pressure broadening can require more than one Lorentzian parameter because of:

  • speed-dependent broadening and shifting;
  • velocity-changing collisions and Dicke narrowing;
  • correlations between state- and velocity-changing collisions;
  • line mixing;
  • quasistatic Stark or van der Waals wings;
  • state-changing collisions; and
  • nonideal density or temperature dependence.

The temperature exponent nn is line and perturber dependent. It is not a universal kinetic-theory constant.

Consider a closed two-level atom whose excited state decays back to its ground state at rate Γ1\Gamma_1, with coherence decay Γ2=Γ1/2+γϕ\Gamma_2=\Gamma_1/2+\gamma_\phi. Define the rotating-frame Hamiltonian by

Hℏ=−Δ∣e⟩⟨e∣+Ω2(∣e⟩⟨g∣+∣g⟩⟨e∣),\frac{H}{\hbar}=-\Delta|e\rangle\langle e| +\frac{\Omega}{2}(|e\rangle\langle g|+|g\rangle\langle e|),

where Ω\Omega is real, Δ=ωL−ω0\Delta=\omega_{\rm L}-\omega_0, and the rotating-wave and Markov approximations apply. The steady excited-state population as the laser frequency is scanned is

ρee(Δ)=s0/21+s0+(Δ/Γ2)2,s0=Ω2Γ1Γ2.\rho_{ee}(\Delta)= \frac{s_0/2}{1+s_0+(\Delta/\Gamma_2)^2}, \qquad s_0=\frac{\Omega^2}{\Gamma_1\Gamma_2}.

Thus the scan has angular-frequency widths

γpow,ω=Γ21+s0,wpow,ω=2Γ21+s0.\gamma_{\rm pow,\omega}=\Gamma_2\sqrt{1+s_0}, \qquad w_{\rm pow,\omega}=2\Gamma_2\sqrt{1+s_0}.

The fluorescence count rate integrated over emitted frequencies is Γ1ρee\Gamma_1\rho_{ee} and has the same scan width. By contrast, a frequency-resolved emission spectrum at fixed laser frequency is a different measurement: even this ideal two-level atom can show a Mollow triplet under strong drive. Do not use the scan-width formula for the separation or width of those emission peaks. See Optical Bloch Equations.

For independent particles with a stationary velocity distribution,

S(Δ)=∫p(v) S2L(Δ−kv;Ω,Γ1,Γ2) dv.S(\Delta)=\int p(v)\, S_{\rm 2L}(\Delta-kv;\Omega,\Gamma_1,\Gamma_2)\,dv.

If p(v)p(v) is Gaussian and all velocity classes experience the same Ω\Omega, Γ1\Gamma_1, and Γ2\Gamma_2, this is a Voigt shape with the power-broadened Lorentzian width above. Spatially varying intensity, velocity-dependent optical pumping, multilevel dark states, or nonequilibrium motion require the corresponding average of the actual response. Power broadening is already part of that response; adding another independent “power width” would count it twice.

Contribution Typical profile or forward model Key caution
transit time depends on beam geometry and trajectory distribution not necessarily a Lorentzian and not an intrinsic level lifetime
finite square interrogation sinc-squared probability envelope Fourier-limited width is a measurement response, not a decay width
Ramsey interrogation central fringe plus side fringes report pulse separation and estimator
Gaussian instrument response convolution; variances add to other Gaussian stages calibrate with a source narrower than the instrument
finite spectrometer slit apparatus-specific line-spread function may be triangular, sinc-like, or asymmetric
laser frequency noise depends on noise spectrum and observation time one stationary linewidth may not exist
unresolved hyperfine or isotope components weighted sum of shifted lines can mimic broadening, asymmetry, or line pulling
opacity and saturation in propagation nonlinear radiative-transfer model transmitted intensity need not preserve the microscopic profile

For laser phase noise and coherence, use Linewidth and Coherence. For finite-time estimators and systematic shifts, use Precision Spectroscopy.

The distinction is operational:

Class Meaning Standard examples Can subensemble selection or refocusing reduce it?
homogeneous every selected member has the same dynamical profile lifetime decay, Markovian pure dephasing, impact-limit collisions not by merely selecting a narrower static class
inhomogeneous members have different static or slowly varying centers Doppler shifts, site disorder, field gradients, slow detuning drift often, if the offset is resolvable or reversible

For a static center distribution p(Ω)p(\Omega) and a center-independent homogeneous profile,

Lens(ω)=∫−∞∞p(Ω)Lhom(ω−Ω) dΩ.L_{\mathrm{ens}}(\omega) = \int_{-\infty}^{\infty} p(\Omega) L_{\mathrm{hom}}(\omega-\Omega) \,d\Omega.

A Gaussian pp and Lorentzian LhomL_{\mathrm{hom}} produce a Voigt profile. However, the same environment can appear inhomogeneous during a short scan and homogeneous during a long coherence interval if its fluctuation time changes relative to the measurement.

The ensemble dephasing time T2∗T_2^* can therefore be shorter than the homogeneous T2T_2. An echo can refocus sufficiently slow reversible offsets, but not population loss or arbitrarily fast dephasing.

The ordinary Voigt model assumes an independent Gaussian distribution of centers and a velocity-independent Lorentzian width and shift. High-resolution gas spectra can resolve violations of those assumptions.

Observed behavior Physical model to consider
width or shift varies across speed classes speed-dependent Voigt
collisions narrow the Doppler core velocity-changing collisions or Dicke narrowing
both effects matter Hartmann–Tran-family profile
neighboring transitions produce coupled response line-mixing model
shoulders or multiple peaks explicit hyperfine, isotope, or dressed-state components

The modified Hartmann–Tran model described by Wcisło et al. (2025) includes effects beyond ordinary Voigt broadening. Extra parameters are useful only when the measurements can constrain them. A structured residual can also come from the instrument, background subtraction, or unresolved lines; its shape does not uniquely identify a microscopic cause.

For a first calculation, check profile normalization and the Gaussian and Lorentzian limiting cases. For a fit, compare the core and wings, vary the fit window, and use independently measured instrument response, temperature, pressure, and drive intensity where available. The fuller discussion is in Line Shapes and Broadening.

A fitted width does not uniquely identify its cause. Gaussian broadening can include instrument resolution and slow frequency drift; Lorentzian broadening can include collisions and pure dephasing. A pressure, temperature, or intensity series can distinguish mechanisms that one spectrum cannot.

Removing instrumental width depends on the profile. Independent Gaussian variances subtract, so Gaussian FWHMs subtract in quadrature. Lorentzian HWHMs subtract linearly. A measured response of another shape should be included through its convolution.

The transmitted intensity is not the absorption coefficient. At large optical depth, exponential attenuation changes the observed profile. Fit the quantity the detector measures, including the propagation model.

An upper state has radiative lifetime τ=30.0\tau=30.0 ns. The lower state is stable and pure dephasing is negligible. Find the angular-frequency FWHM, ordinary-frequency FWHM, and Lorentzian HWHM in MHz.

Solution

The population-decay rate is

Γu=1τ=3.333×107 s−1.\Gamma_u = \frac{1}{\tau} = 3.333\times10^7\ \mathrm{s^{-1}}.

In the stated limit, this is the angular-frequency FWHM:

wnat,ω=3.333×107 rad s−1.w_{\mathrm{nat},\omega} = 3.333\times10^7\ \mathrm{rad\,s^{-1}}.

The ordinary-frequency FWHM is

Δνnat=Γu2π≃5.31 MHz.\Delta\nu_{\mathrm{nat}} = \frac{\Gamma_u}{2\pi} \simeq 5.31\ \mathrm{MHz}.

The ordinary-frequency HWHM is half of that:

γL,ν≃2.65 MHz.\gamma_{\mathrm L,\nu} \simeq 2.65\ \mathrm{MHz}.

Estimate the ordinary-frequency Doppler FWHM of a 780780 nm transition in a 87Rb^{87}\mathrm{Rb} vapor at 300300 K. Use m=87 um=87\,u, u=1.66054×10−27u=1.66054\times10^{-27} kg, and kB=1.380649×10−23 J K−1k_{\mathrm B}=1.380649\times10^{-23}\ \mathrm{J\,K^{-1}}.

Solution

The center frequency is

ν0=cλ0≃3.8435×1014 Hz,\nu_0 = \frac{c}{\lambda_0} \simeq 3.8435\times10^{14}\ \mathrm{Hz},

and the mass is

m=87u≃1.4447×10−25 kg.m = 87u \simeq 1.4447\times10^{-25}\ \mathrm{kg}.

Use

ΔνD=2ν02kBTln⁡2mc2.\Delta\nu_{\mathrm D} = 2\nu_0 \sqrt{ \frac{2k_{\mathrm B}T\ln2} {mc^2} }.

Substitution gives

ΔνD≃5.11×108 Hz,\Delta\nu_{\mathrm D} \simeq 5.11\times10^8\ \mathrm{Hz},

or about

511 MHz.511\ \mathrm{MHz}.

This is much broader than the few-megahertz natural width of a typical alkali resonance, which is why room-temperature vapor spectroscopy often uses Doppler-free methods to resolve hyperfine-scale structure.

At 296296 K, a molecular line has air-broadened HWHM coefficient γair=0.070 cm−1 atm−1\gamma_{\mathrm{air}}=0.070\ \mathrm{cm^{-1}\,atm^{-1}} and self-broadened coefficient γself=0.30 cm−1 atm−1\gamma_{\mathrm{self}}=0.30\ \mathrm{cm^{-1}\,atm^{-1}}. The total pressure is 0.800.80 atm and the absorber partial pressure is 0.0200.020 atm. Neglect temperature correction. Find the Lorentzian HWHM and FWHM.

Solution

The air partial pressure is

pair=0.80−0.020=0.780 atm.p_{\mathrm{air}} = 0.80-0.020 = 0.780\ \mathrm{atm}.

The HWHM is

γL=γairpair+γselfpself=(0.070)(0.780)+(0.30)(0.020)=0.0606 cm−1.\begin{aligned} \gamma_{\mathrm L} &= \gamma_{\mathrm{air}}p_{\mathrm{air}} + \gamma_{\mathrm{self}}p_{\mathrm{self}} \\ &= (0.070)(0.780) + (0.30)(0.020) \\ &= 0.0606\ \mathrm{cm^{-1}}. \end{aligned}

Therefore,

wL=2γL=0.1212 cm−1.w_{\mathrm L} = 2\gamma_{\mathrm L} = 0.1212\ \mathrm{cm^{-1}}.

A pressure shift coefficient, if supplied, would modify the center rather than this width.

A closed atom has Γ1=2π×6 MHz\Gamma_1=2\pi\times6\ {\rm MHz}, no pure dephasing, and Rabi frequency Ω=2π×3 MHz\Omega=2\pi\times3\ {\rm MHz}. Find the ordinary-frequency FWHM of its steady fluorescence signal as the laser frequency is scanned.

Solution

Here Γ2=Γ1/2\Gamma_2=\Gamma_1/2 and s0=2Ω2/Γ12=1/2s_0=2\Omega^2/\Gamma_1^2=1/2. The FWHM is

Δν=Γ12π1+s0=61.5 MHz≃7.35 MHz.\Delta\nu=\frac{\Gamma_1}{2\pi}\sqrt{1+s_0} =6\sqrt{1.5}\ {\rm MHz}\simeq7.35\ {\rm MHz}.

This is a scan of the integrated fluorescence signal. It is not a statement about the frequency-resolved spectrum emitted at fixed drive.

A source has spectral power density Pν(ν)P_\nu(\nu) and passes through a sample with transmission T(ν)=e−α(ν)ℓT(\nu)=e^{-\alpha(\nu)\ell}. Write both the transmitted spectral power density and the transmission when the horizontal axis is changed to vacuum wavelength.

Solution Pλ,out(λ)=Pν,in(c/λ)T(c/λ)cλ2.P_{\lambda,\rm out}(\lambda)= P_{\nu,\rm in}(c/\lambda)T(c/\lambda)\frac{c}{\lambda^2}.

The incident density has the same factor c/λ2c/\lambda^2. Dividing the two densities gives Tλ(λ)=Tν(c/λ)T_\lambda(\lambda)=T_\nu(c/\lambda). Therefore αλ(λ)=αν(c/λ)\alpha_\lambda(\lambda)=\alpha_\nu(c/\lambda) also has no Jacobian. Power per wavelength interval transforms as a density; the attenuation of a specified monochromatic beam does not.

  • Demtröder, Wolfgang. Laser Spectroscopy 1: Basic Principles, 5th ed. Springer (2014). doi:10.1007/978-3-642-53859-9. Background for the lifetime, Doppler, collision, and saturation formulas.
  • HITRAN. “Definitions and Units”. Online documentation, accessed 2026-10-14. Defines molecular line strengths, HWHM pressure coefficients, and temperature conventions.
  • Martin, William C., Wolfgang L. Wiese, and Alexander Kramida. “Spectral Line Shapes, Widths, and Shifts”, in Atomic Spectroscopy: An Introduction. NIST, online reference, accessed 2026-10-14.
  • NIST Digital Library of Mathematical Functions. “§7.19: Voigt Functions”, accessed 2026-10-14.
  • Olivero, J. J., and R. L. Longbothum. “Empirical fits to the Voigt line width: A brief review.” Journal of Quantitative Spectroscopy and Radiative Transfer 17, 233–236 (1977). doi:10.1016/0022-4073(77)90161-3.
  • Wcisło, P., et al. “New beyond-Voigt line-shape profile recommended for the HITRAN database.” Journal of Quantitative Spectroscopy and Radiative Transfer 347, 109596 (2025). doi:10.1016/j.jqsrt.2025.109596; author manuscript.